Radiation Therapy
Roos Electron Ionization
Chamber, 0.35 cm 3 , Waterproof
Model 30-332
•  Designed as a standard chamber
for electron dosimetry
•  Wide guard ring excludes
perturbation effects, even at low
electron energies
•  Negligible polarity effect, < 0.5% at
10 MeV
•  Energy dependence is only
influenced by the stopping power
correction, a type dependent
correction is not necessary
•  Completely Waterproof
•  Ideal for use in water phantoms or
suitable solid state phantoms
•  Open volume, vented
The Roos chamber, a development from Dr. Roos of PTB-Braunschweig, is used as a
standard chamber for electron dosimetry. This chamber has a very wide guard ring to
exclude any perturbation effect, even at low electron energies. The polarity effect is
negligible. Energy dependence is only influenced by the stopping power correction, a
type dependent correction is not necessary. The chamber is waterproof and vented through
the connection cable. Roos includes a 3.5 ft (1.08 m) cable and BNC Triax connector.
Specifications
Volume 0.35 cm 3
Response 1 • 10 -8 C/Gy
Leakage ± 4 10 -15 A
Polarizing voltage 100 V recommended, maximum 400 V
Cable length 3.5 ft (1.08 m)
Cable leakage 3.5 10 -12 C/(Gy cm)
Wall material Acrylic (C 5 H 8 O 2 )
Wall density 1.19 gm/cm 3
Wall thickness 1.0 mm
Area density 119 mg/cm 2
Electrode Acrylic, graphite coated, 15 mm Ø
Guard ring 4 mm wide
Nominal useful range 2 to 25 MeV
Range of temperature 10º to 40ºC
Range of relative humidity 10% to 80%
Ion collection time
100 V: 0.37 ms
200 V: 0.13 ms
300 V: 0.07 ms
Saturation behavior
Saturation
behavior
Polarizing
voltage
99.0%
saturation
99.5%
saturation
Maximum dose rate at
continuous irradiation
100 V
2.6 Gy/s
1.3 Gy/s
200 V
11.0 Gy/s
5.2 Gy/s
400 V
42.0 Gy/s
21.0 Gy/s
100 V
0.5 mGy
0.2 mGy
Maximum dose rate
per irradiation pulse
200 V
0.9 mGy
0.5 mGy
400 V
1.9 mGy
0.9 mGy
See next page for more specifications.
 
Specifications (continued)
Diagram
Ø 4 4
Ø 15 .6 (el ec trod e)
Ø 1 6
4 (g u a rd ri ng )
Approximate dimensions in mm.
Drawing not to scale.
Ø 2 4
5.8
2
1
Directional dependence
The deviation of the response following tilting of the chamber by up to 10º at 6 and 20 MeV,
at the dose maximum in water, is less than 0.1%.
Optional accessories
Check Source, 90 Sr, 20 MBq (541 µCi) (Model 30-658)
Available model(s)
30-332 Roos Electron Ionization Chamber, 0.35 cm 3 , Waterproof,
includes BNC Triax connector
Other types of triaxial cable connectors available