Radiation Therapy
Roos
™
Electron Ionization
Chamber, 0.35 cm
3
, Waterproof
Model 30-332
•
Designed as a standard chamber
for electron dosimetry
•
Wide guard ring excludes
perturbation effects, even at low
electron energies
•
Negligible polarity effect, < 0.5% at
10 MeV
•
Energy dependence is only
influenced by the stopping power
correction, a type dependent
correction is not necessary
•
Completely Waterproof
•
Ideal for use in water phantoms or
suitable solid state phantoms
•
Open volume, vented
The Roos chamber, a development from Dr. Roos of PTB-Braunschweig, is used as a
standard chamber for electron dosimetry. This chamber has a very wide guard ring to
exclude any perturbation effect, even at low electron energies. The polarity effect is
negligible. Energy dependence is only influenced by the stopping power correction, a
type dependent correction is not necessary. The chamber is waterproof and vented through
the connection cable. Roos includes a 3.5 ft (1.08 m) cable and BNC Triax connector.
Specifications
Volume
0.35 cm
3
Response
1 • 10
-8
C/Gy
Leakage
± 4 10
-15
A
Polarizing voltage
100 V recommended, maximum 400 V
Cable length
3.5 ft (1.08 m)
Cable leakage
3.5 10
-12
C/(Gy cm)
Wall material
Acrylic (C
5
H
8
O
2
)
Wall density
1.19 gm/cm
3
Wall thickness
1.0 mm
Area density
119 mg/cm
2
Electrode
Acrylic, graphite coated, 15 mm Ø
Guard ring
4 mm wide
Nominal useful range
2 to 25 MeV
Range of temperature
10º to 40ºC
Range of relative humidity
10% to 80%
Ion collection time
100 V:
0.37 ms
200 V:
0.13 ms
300 V:
0.07 ms
Saturation behavior
Saturation
behavior
Polarizing
voltage
99.0%
saturation
99.5%
saturation
Maximum dose rate at
continuous irradiation
100 V
2.6 Gy/s
1.3 Gy/s
200 V
11.0 Gy/s
5.2 Gy/s
400 V
42.0 Gy/s
21.0 Gy/s
100 V
0.5 mGy
0.2 mGy
Maximum dose rate
per irradiation pulse
200 V
0.9 mGy
0.5 mGy
400 V
1.9 mGy
0.9 mGy
See next page for more specifications.
Specifications
(continued)
Diagram
Ø 4 4
Ø 15 .6 (el ec trod e)
Ø 1 6
4 (g u a rd ri ng )
Approximate dimensions in mm.
Drawing not to scale.
Ø 2 4
5.8
2
1
Directional dependence
The deviation of the response following tilting of the chamber by up to 10º at 6 and 20 MeV,
at the dose maximum in water, is less than 0.1%.
Optional accessories
Check Source,
90
Sr, 20 MBq (541 µCi) (Model 30-658)
Available model(s)
30-332
Roos Electron Ionization Chamber, 0.35 cm
3
, Waterproof,
includes BNC Triax connector
Other types of triaxial cable connectors available